TEACHING IN THE LABORATORY
k Hostounsk
11 The Stentor Institute, Praha-Západ 2 Institute of Microbiology, Academy of Sciences of the Czech Republic, Praha, Czechoslovakia
Address for reprint requests and other correspondence: R. Pelc, Institute of Microbiology, Academy of Sciences of the Czech Republic, K V
zkumn
m ústavüm 1083, CZ-14220 Praha 4-krc, Czechoslovakia (e-mail: rpelc{at}biomed.cas.cz and radekopelc{at}seh.oxon.org)
| Abstract |
|---|
|
|
|---|
Key words: three-dimensional impression and differential imaging; depth perception; relief diaphragm; shape-from-shading; visual cues
| Introduction |
|---|
|
|
|---|
In the biology practical class, the optical microscope represents the most common "interface" between the macroscopic and little-known microscopic world. Optical artifacts stemming from an incorrectly adjusted microscope illumination are well known, e.g., the diffraction circles when the condenser aperture is closed too much. However, various artifacts still persist in the image even at the optimal (Köhler) setting (6, 11). These include, e.g., the disturbing "halo" artifact in the images obtained under the phase-contrast microscope (1, 6, 13, 19). A refined version, apodized phase contrast (9), introduced for the first time by Nikon (e.g., United States Patent No. 6317261), can partially eliminate the halo.
Here, we demonstrate that some of these issues can be easily addressed in a laboratory classroom by using, as an example, a simple biological specimen and a simple reference specimen, both of which can be easily prepared and examined under two different illumination modes commonly used in optical microscopy, namely, phase-contrast and off-axis illumination. The aim is to expand the scope of approaches that can be used to demonstrate various optical illusions (6–8, 14) and to relate the outcome of observation to the complexity of both the specimen and experimental apparatus.
| EXPERIMENTAL PROCEDURES |
|---|
|
|
|---|
An upright laboratory microscope [model DN45–BH51, Lambda Praha (formerly Meopta Czechoslovakia)] was used. The (negative) phase-contrast images were taken with Plan 10Ph/0.25 (radula) or Plan 20Ph/0.40 [electron microscopical (EM) grid imprint] objectives (planachromatic); a condenser normally supplied with the model DN45-BH47Ph microscope was used (Lambda Praha, numerical aperture 1.25). All other images were acquired with 10/0.25 (radula) or 20/0.45 (EM grid imprint) objectives (achromatic) and a shifting relief diaphragm condenser (Abbé type, numerical aperture 1.2, model RCH-0128), with the relief diaphragm either engaged (off–axis illumination) or disengaged (axial, i.e., ordinary bright-field illumination). The condenser itself is shown in Fig. 1. All photomicrographs were taken with Minolta X300s or Praktica VLC3 camera on Fuji Superia 100 color print film; a photographic eyepiece (FU x4, Lambda Praha) and a green filter were used.
|
INSTRUCTIONS.
Cut a "D"–shaped piece from a black cardboard ("relief diaphragm") and place it as close as possible to condenser's iris diaphragm (front focal plane). Most condensers in upright microscopes are fitted with a filter holder placed just under the iris diaphragm. The relief diaphragm can be placed into this holder, which can be usually swung out of the optical axis. When viewing the specimen, try to experiment with the position and orientation of the relief diaphragm until the most satisfactory image is obtained. The most natural orientation of the diaphragm is the one providing an apparent illumination angle from the "north" (or northwest/northeast), which is reminiscent of overhead (sun) illumination (see RESULTS AND DISCUSSION for details).
Various other shapes of the relief diaphragm can be used as well, e.g., with its edge slightly concave, as in the present study or elsewhere (4, 5), or convex. A positive lens (instead of the relief diaphragm) placed in the above-mentioned filter holder slightly swung out from the condenser's optical axis represents a less efficient but usable way of producing off–axis illumination (18).
The relief diaphragm does not necessarily have to be inserted into (or close to) the front focal plane of the condenser. Other optical planes that are conjugate with it may be modified instead (6) or in addition (15). One of these options is the objective back focal plane, which is usually accessible in low-magnification objectives (1). In this way, a pseudorelief [quasi-three-dimensional (3-D)] effect is achieved in a manner similar to the more complex Hoffman modulation contrast (3). These various possibilities have been described in detail elsewhere (1, 5, 6, 15).
Note that it is critical to bring the relief diaphragm as close as possible to the condenser's front focal plane or a plane conjugate with it (6). This guarantees not only a strong relief (3-D) impression in the images but a uniform illumination of the viewing field as well. Failure to comply with this basic requirement can easily lead to rather disappointing results in that only a small portion of the image is of satisfactory quality (10), a factor that is also of importance in macroscopic schlieren imaging (15). In some condensers, mechanical constraints may not permit the relief diaphragm to be placed close enough to the iris diaphragm. In that case, an accessory lens, which is, in some condensers, already built in immediately under the iris diaphragm, may be utilized to bring the two diaphragms optically closer together, as in the present design (5) (Fig. 1).
Specimen Preparation
The biological specimen. The biological specimen (radula, a translucent chitinous ribbon in the mouth of certain mollusks) was isolated from the Spanish slug [Arion lusitanicus Mabille 1868 (Stylomatophora, Arionidae)], which was obtained locally.
INSTRUCTIONS.
Let the slug stretch out. Cut off the front part (
1/4 of the slug) with a scalpel or scissors, and submerge it overnight into a concentrated solution of NaOH [
10% (wt/vol)].1
Expose the radula by removing the remnants of the soft tissues, and then rinse the specimen in water, dry it, place it on a microslide, and cover it with a coverslip. It is recommended to attach the coverslip's edges to the microslide, e.g., with wax or Canada balsam, to keep the radula in place. In the adult slug we used (
10 cm long), the radula was a yellowish flat object ("scale")
4 mm long and
2 mm wide. Other slug or snail species would also be suitable. However, the animal should be at least
2 in. (5 cm) long. Otherwise, the radula may be too small and difficult to locate and handle, and its relief structure may not be prominent enough to enable proper visualization under off-axis illumination.
The reference specimen. The reference specimen was prepared from gelatin.
INSTRUCTIONS.
Make a few milliliters of an aqueous gelatin solution [
5–10% (wt/vol),
50°C] in either a beaker placed in a water bath or a test tube kept warm on a laboratory burner (gelatin does not dissolve in cold water). Place a small gelatin drop on a microslide and let it spread out. Place an EM grid on the (still liquid) gelatin layer. A copper grid was used here of mesh 300 (meaning 300 "windows"/in.), a window width of
54 µm, and a thickness of
20 µm. Place the microslide into a refrigerator (for
10 min) for the gelatin to harden. Remove the EM grid from the gelatin with a sharp pair of tweezers or a needle. At an appropriate angle, an imprint in the gelatin of the EM grid can be seen with a naked eye.
Examination of the Specimens
In both the radula and EM grid imprint specimens, the axial and off–axis illumination images were taken on exactly the same portion of the specimen while the phase-contrast images were acquired very close to that area (Fig. 2). In the radula specimen, all images were taken within 450 µm from its longitudinal axis and on the same side from that axis. Further details can be found in the text below and in Fig. 2.
|
| RESULTS AND DISCUSSION |
|---|
|
|
|---|
The halo artifact in the phase-contrast images is profound in both preparations and is particularly disturbing in the radula specimen, as the density of prominent edges is very high. Indeed, the radula is designed for rasping food and has to be equipped with many indentations (teeth or denticles). The radula examined in the present study contained
85 longitudinal and 125 transversal rows of teeth. The biggest ones are in radula's center, with their spacing being
40 µm along the radula's longitudinal axis (=longitudinal row 0) by
25 µm. At the same time, the performance of the phase contrast is poor on too-thick specimens. This is partly due to the depth of field, which is smaller in the phase-contrast illumination mode compared with the off-axis illumination one, i.e., not due to the halo artifact alone. For this reason, the phase-contrast image of the radula was deliberately taken on a somewhat thinner portion of the specimen; note that the teeth are slightly smaller (Fig. 2).
The halo artifact, accompanied by the so-called shade-off (or shading-off) artifact, is brought about by sufficiently large optical thickness gradients (13). The halo is absent, e.g., on the edges of very thin cell extensions (6, 13). One way of reducing it is to place the object of interest into a solution of matching refractive index (13), but this approach cannot be applied to every object (e.g., highly refractive granules inside cells).
In this context, it is important to note that the human brain treats the visual scene (the microscopic image) as if it was acquired under the incident overhead (sun) illumination common in the macroscopic world, where most objects of interest are nontranslucent (with the rare exceptions being, e.g., a translucent tree leaf, a spider web, or a stained glass window in a church), with darker and brighter areas (shading) on their surface, depending on the direction of illumination and the object's shape. The shading patterns are then interpreted as projections and depressions of the object's surface (12). In optical microscopy, this effect can be used, e.g., when choosing the most convenient apparent direction of illumination (20).
These factors alone speak in favor of the off–axis, or oblique, illumination mode over the phase-contrast mode even if the halo artifact is absent in phase-contrast images. The reason is that, of these two imaging modes, only the former is capable of mimicking the above-mentioned shading patterns encountered on the surfaces of macroscopic, mostly nontranslucent objects, by producing so-called schlieren images (15). The human visual system is thus presented with a more comprehensible type of microscopic image in that the images are enhanced with so-called visual cues to depth structure (6). This essentially enables "on-line" 3-D reconstruction, directly in the experimentalist's brain.
It is true that the shapes interpreted from the schlieren images do not necessarily represent a real physical profile; the luminance distribution (shading profile) in the image approximately translates to the distribution of the refractive index gradients (slopes angles) (15). These gradients correspond to local variations in optical rather than physical thickness. However, optical and physical thickness profiles do coincide in an object with a uniform refractive index distribution, with this condition being more or less satisfied in the radula or gelatin specimens.
The shading profiles are particularly strongly modified at the object's edges (contours) by the off-axis (or generally schlieren) illumination. Overall image interpretation has been shown to be affected by their shape (12) and completeness (16). Involuntary saccadic (jumping) eye movements (6) may play a role in these phenomena by highlighting the edges (2).
In conclusion, the authors believe that this type of practical demonstration, suitable for any biology college classroom, would enable the students to 1) prepare both a biological and reference specimen capable of unmasking some weak points of a technique that is now a commonplace in the biological laboratory, the phase-contrast microscopy [1953 Nobel Prize to Frits Zernike (19)]; 2) convert a laboratory microscope to the off–axis illumination mode capable of producing, in this particular case, better results than phase contrast itself; 3) compare visual impressions from the two image types and understand some of the physiological aspects of depth perception; and 4) realize the need for a careful choice of an apparatus, with the "bottom line" being that more complex methods may not necessarily yield better results. The overall aim of the present article is to induce in students critical thinking about the capabilities of laboratory equipment in general and about data interpretation.
| GRANTS |
|---|
|
|
|---|
| Acknowledgments |
|---|
and R. Pelc, respectively. | Footnotes |
|---|
| REFERENCES |
|---|
|
|
|---|
Z. Osv
tlovací soustava pro mikroskopické pozorování prühledn
ch neabsorbujících objektü. Czech patent CZ4925 (utility model), 1996.
Z, Pelc R. An efficient way of high-contrast, quasi–3D cellular imaging: off–axis illumination. J Biochem Biophys Meth 68: 23–30, 2006.[CrossRef][Web of Science][Medline]
ek J, Reischig J. Transmitted-light microscopy for biology: a physicist's point of view. Parts I and II. Proc Roy Microsc Soc 33: 121–127 and 196–205, 1998.
i
ka Z, Hostounsk
Z. RCH–a ARC-mikroskopie–nové kontrastní metody. Jemná Mechanika a Optika 7–8: 199–201, 2004.
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||